Common Deployment Settings

Common Settings & Battery Life

The table below describes the most common settings used with the TCM Tilt Current Meters.

Configuration Temperature (minute) Burst Interval (minute) Burst Rate (Hz) Burst Duration (seconds) Start Time Stop Time Notes
Typical 1 1 8 20 On Next Minute Until Stopped Suitable for most environments, high temporal resolution with moderate filtering. 12 to 14 month typical battery life
Swells 2 2 8 45 On Next Minute Until Stopped Recommended for locations with swells with a period of up to 10 seconds. 10 to 12 month typical battery life
Energetic / Turbulence 1 1 16 30 On Next Hour Until Stopped Recommended for areas near surf zones or other very turbulent locations. 5 to 6 month typical battery life. Increase burst duration to 30 seconds.
Continuous Recording 1 second 1 second 16 1 Immediately Until Stopped Data logging at 16Hz continuously. 3 month typical battery life
Burst Recording 5 5 16 60 On Next Quarter Hour Until Stopped Data logging at 16 Hz, 1 minute seconds ON, 4 minutes OFF. 12 month typical battery life.